• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON CR X3 TIBIAL INSERT; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON CR X3 TIBIAL INSERT; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO Back to Search Results
Model Number 5530-G-309
Medical Device Problem Codes Degraded (1153); Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Joint Laxity (4526); Insufficient Information (4580)
Date of Event 02/01/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.Device not returned to the manufacturer.
 
Event or Problem Description
The following was reported: polyethylene insert was revised after 5 years post surgery.
 
Event or Problem Description
The following was reported: polyethylene insert was revised after 5 years post surgery.
 
Additional Manufacturer Narrative
Reported event: an event regarding wear involving a triathlon insert was reported.The event was confirmed via evaluation of the returned device and clinician review of the provided medical records.Method & results: -product evaluation and results: visual inspection: visual inspection of the returned device indicated that the insert is worn.The damage present on the posterior portion was consistent with delamination and material loss.The damage on the articulating surface of the insert is consistent with burnishing, scratching, and third body indentations; which are common damage modes of uhmwpe.The yellow discoloration observed on the tibial insert is consistent with the absorption of synovial fluid by the device.-clinician review: a review of the provided medical records by a clinical consultant indicated: "this case concerns a female patient who underwent a cemented primary cruciate retaining triathlon total knee arthroplasty.Approximately five years later the patient required revision for polyethylene wear and anterior tibial subluxation.I can confirm that the primary procedure took place and that a revision procedure took place since i can see a post revision x-ray and the date of explant was noted in the summary.However i do not have any further documentation of the revision procedure.I cannot determine the root cause of this event with certainty.The root causes of polyethylene wear with anterior subluxation are multifactorial.These include surgical technique factors, patient factors including activity level and bmi and implant factors.With a posterior cruciate ligament retaining implant if the posterior cruciate ligament is left relatively tight, that creates a relatively tight flexion space which can result in wear of the polyethylene posteriorly.Ligament balancing and restoration of kinematics also impact the wear of polyethylene." -product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.-complaint history review: there have been no other similar events for the lot referenced.Conclusion: it was reported that the patient was revised.Visual inspection of the returned device indicated that the insert is worn.The damage present on the posterior portion was consistent with delamination and material loss.The damage on the articulating surface of the insert is consistent with burnishing, scratching, and third body indentations; which are common damage modes of uhmwpe.The yellow discoloration observed on the tibial insert is consistent with the absorption of synovial fluid by the device.A review of the provided medical records by a clinical consultant indicated the following: "this case concerns a female patient who underwent a cemented primary cruciate retaining triathlon total knee arthroplasty.Approximately five years later the patient required revision for polyethylene wear and anterior tibial subluxation.I cannot determine the root cause of this event with certainty.The root causes of polyethylene wear with anterior subluxation are multifactorial.These include surgical technique factors, patient factors including activity level and bmi and implant factors.With a posterior cruciate ligament retaining implant if the posterior cruciate ligament is left relatively tight, that creates a relatively tight flexion space which can result in wear of the polyethylene posteriorly.Ligament balancing and restoration of kinematics also impact the wear of polyethylene." no further investigation for this event is possible at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.Correction - d6a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIATHLON CR X3 TIBIAL INSERT
Common Device Name
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
arokiya raj
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key16450866
Report Number0002249697-2023-00197
Device Sequence Number20302333
Product Code MBH
UDI-Device Identifier07613327050431
UDI-Public07613327050431
Combination Product (Y/N)N
Initial Reporter CountryEI
PMA/510(K) Number
K051146
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2023
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/31/2020
Device Model Number5530-G-309
Device Catalogue Number5530-G-309
Device Lot Number2325EM
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/01/2023
Supplement Date Received by Manufacturer07/11/2023
Initial Report FDA Received Date02/28/2023
Supplement Report FDA Received Date07/27/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/29/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexFemale
-
-