It was reported that during a recanalization procedure for thrombosed arteriovenous graft, the device allegedly got caught on to a valve and tore the vein open.It was further reported that, the patient allegedly had a hole in vessel wall and experienced extravasation caused by the device.Reportedly, vein reconstruction was performed and some covered stents were used to help cover the extravasation.The procedure was completed using another device.The patient is stable now.
|
|
The medical device manufacturer and manufacturing location for the straub product was selected as unknown due to system limitations.The correct medical device manufacturer and manufacturing location are straub medical us.As the lot number for the device was provided, a review of the device history records will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.(expiration date: 11/2024).The device not returned.
|