| Model Number |
MQK1610 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/31/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The manufacturing location was selected as unknown due to system limitations.The manufacturing location for the this product is bd-santo domingo.As the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.Expiry date: 09/2025.Device pending return.
|
| |
|
Event or Problem Description
|
|
It was reported that during an ultrasound-guided biopsy procedure, the cap of the introducer needle allegedly came off leaving the needle within the introducer.Reportedly, the needle was retrieved out of the introducer and proceeded with the procedure.The procedure was completed using same device there was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: the manufacturing location was selected as unknown due to system limitations.The manufacturing location for the this product is bd-santo domingo.H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: the physical device was not returned for evaluation.No photos were provided for review.Therefore, the investigation is inconclusive for the reported failure as no objective evidence was provided for review.A definitive root cause for the alleged break issue could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiry date: 09/2025).H11: section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that during an ultrasound-guided biopsy procedure, the cap of the introducer needle allegedly came off leaving the needle within the introducer.Reportedly, the needle was retrieved out of the introducer and proceeded with the procedure.The procedure was completed using same device there was no reported patient injury.
|
| |
|