MAUDE Adverse Event Report: ARROW INTERNATIONAL LLC ARROW HEMODIALYSIS SET: 2-LUMEN 14 FR X 20 CM; CATHETER HEMODIALYTSIS NON IMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL LLC ARROW HEMODIALYSIS SET: 2-LUMEN 14 FR X 20 CM; CATHETER HEMODIALYTSIS NON IMP Back to Search Results
Model Number IPN030157
Medical Device Problem Code Air/Gas in Device (4062)
Health Effect - Clinical Code Air Embolism (1697)
Date of Event 01/01/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
Customer complaint is reported as: "a couple of weeks ago, we had a patient suffer an air embolism just after standing up with a renal replacement circuit connected to a 20cm arrowgard line.Air was seen to be being dragged from the patient round the circuit.The line was positioned very close to a vygon central line.The insertion site was neck.Both were removed & examined.There was no evidence of air leak when they were injected with air after being submersed in water and no evidence of a leak when flushed with saline.The easiest explanation is perhaps that the line was moved during mobilisation but the people who were there say that this didn't happen.The patient did have a brief period of deterioration but recovered quite quickly with no ill effects and has been discharged from icu.".
 
Event or Problem Description
Customer complaint is reported as: "a couple of weeks ago, we had a patient suffer an air embolism just after standing up with a renal replacement circuit connected to a 20cm arrowgard line.Air was seen to be being dragged from the patient round the circuit.The line was positioned very close to a vygon central line.The insertion site was neck.Both were removed & examined.There was no evidence of air leak when they were injected with air after being submersed in water and no evidence of a leak when flushed with saline.The easiest explanation is perhaps that the line was moved during mobilisation but the people who were there say that this didn't happen.The patient did have a brief period of deterioration but recovered quite quickly with no ill effects and has been discharged from icu".
 
Additional Manufacturer Narrative
(b)(4), the customer returned one, 2-lumen hemodialysis catheter.Signs of use in the form of biological material were observed.Visual analysis did not reveal any defects or anomalies.The catheter length from the juncture hub to the distal tip measured 217mm, which is within the specification limits of 207mm-227mm per the catheter product drawing.The catheter body outer diameter measured 4.69mm, which is within the specification limits of 4.62mm-4.72mm per the catheter extrusion product drawing.The catheter was assembled and functionally tested per the instructions for use (ifu).The ifu provided with this kit states , "prepare catheter for insertion by flushing each lumen and clamping or attaching injection caps to appropriate extension lines".The catheter was flushed with a lab inventory syringe filled with water; no leaks or blockages were detected.Amrq-000162 states, "there shall be no liquid leakage in the form of a falling drop of water at 300-320 kpa (43.5-46.4 psi) for 30 sec when tested per bs en iso-10555-1 annex c".Both luer hubs of the catheter were connected to the leak tester.With the distal end occluded, the catheter was pressurized to 300kpa for 30 seconds.No leaks detected.A manual tug test confirmed both the arterial and venous luer hubs were securely attached to their extension lines.The customer did not provide a lot number; therefore, a device history record review was performed based upon a lot number from the sales history data of the customer.No relevant findings were identified.The ifu provided with the kit informs the user, "practitioners must be aware of complications associated with central vein catheters including cardiac tamponade secondary to vessel wall, atrial or ventricular perforation, pleural and mediastinal injuries, air embolism, catheter embolism, thoracic duct laceration, bacteremia, septicemia , thrombosis, inadvertent arterial puncture, nerve damage, hematoma, hemorrhage, dysrhythmias, brachial plexus injury, cardiac arrhythmia, exsanguination, hemothorax, luminal thrombosis, pneumothorax , subcutaneous hematoma, and retroperitoneal bleed".The report of an air embolism could not be confirmed as part of this complaint investigation.The returned catheter met all relevant visual, dimensional, and functional requirements including leak testing performed per bs en iso-10555-1.A device history record review was performed based on a potential lot identified from a sales history review.No relevant findings were identified to suggest a manufacturing related issue.Based on the customer report and the sample received, no problem was found with the returned device.Teleflex will continue to monitor and trend for reports of this nature.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARROW HEMODIALYSIS SET: 2-LUMEN 14 FR X 20 CM
Common Device Name
CATHETER HEMODIALYTSIS NON IMP
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL CR, A.S.
jamska 2359/47
zdar nad sazavou 591 0 1
EZ   591 01
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key16452398
Report Number3006425876-2023-00195
Device Sequence Number20216479
Product Code MPB
UDI-Device Identifier10801902009217
UDI-Public10801902009217
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K993933
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/31/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Model NumberIPN030157
Device Catalogue NumberCS-25142-F
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/31/2023
Supplement Date Received by Manufacturer03/21/2023
Initial Report FDA Received Date02/28/2023
Supplement Report FDA Received Date03/22/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
VYGON CENTRAL LINE; VYGON CENTRAL LINE
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-