| Model Number |
MX-410445 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.Reason for reoperation: rupture.
|
| |
|
Event or Problem Description
|
|
Healthcare reported right side rupture.Device has been explanted.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported, right side rupture.Device has been explanted.
|
| |
|
Search Alerts/Recalls
|