| Catalog Number |
08828679190 |
| Medical Device Problem Codes |
Low Test Results (2458); Non Reproducible Results (4029)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/02/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The initial reporter stated they received discrepant results for one patient sample tested with the elecsys brahms pct assay on a cobas e 801 analytical unit.The sample initially resulted in a pct value of 0.16 ng/ml.The result was reported outside of the laboratory and questioned.The sample was repeated resulting in a pct value of 28.9 ng/ml.A new sample draw of the patient resulted in a pct value of 27.3 ng/ml.
|
| |
|
Additional Manufacturer Narrative
|
|
The last calibration performed on (b)(6) 2023, and the quality controls results do not indicate a general reagent issue.Upon review of the alarm trace, no general instrument issue was detected.It was determined that the tip of the sample probe had a crystal, indicating inadequate daily sample probe manual cleaning by the operator.The crystals could be contaminating the aspirated sample but would not cause the observed event.The investigation could not identify a product problem.The cause of the event could not be determined.
|
| |
|