Model Number 26-1221
Medical Device Problem Code Fail-Safe Did Not Operate (4046)
Health Effect - Clinical Code Brain Injury (2219)
Date of Event 02/08/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Event or Problem Description
A physician reported disengagement failure of a disposable perforator (id 261221) during surgery and dural damage was observed.It occurred at making the first burr hole at the tent.Hemostasis and dural suturing were performed.The drill used was anspach.According to information provided, it is unknown if the drill was electric or pneumatic, or if the angle of approach perpendicular as the ifu (instructions for use).It is also unknown if the perpendicular approach maintained through the drilling process or was there any rocking motion included or if there was a constant downward pressure.
 
Event or Problem Description
N/a.
 
Additional Manufacturer Narrative
Perforator (id 261221) has not been returned for evaluation - device history record (dhr)- there is no indication that the production process may have contributed to this complaint.All test results passed procedural specifications.Failure analysis - visual inspection utilizing unaided eye and was heavily soiled with organic material, but no other anomalies were observed.Spring test attempted, unit had a frozen inner / outer drill that was cleaned and had organic material removed.Once cleaned, the unit passed the spring test as designed.Unit successfully drilled 5 holes with no issues therefore, the complaint condition could not be verified.The root cause is undetermined and was unable to be confirmed in the complaint evaluation.Product was received for analysis and the investigation could not confirm the complaint.Possible root cause: ¿user misuse¿ or ¿drill allows to be set incorrectly¿.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CODMAN DISPOS PERFORATOR
Common Device Name
DISPOSABLE PERFORATORS
Manufacturer (Section D)
RAYNHAM
325 paramount drive
325 paramount drive
raynham MA
Manufacturer (Section G)
RAYNHAM
325 paramount drive
raynham MA
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key16455684
Report Number3014334038-2023-00021
Device Sequence Number12374957
Product Code HBF
UDI-Device Identifier10886704005100
UDI-Public10886704005100
Combination Product (Y/N)N
PMA/510(K) Number
K791101
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/04/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/31/2024
Device Model Number26-1221
Device Catalogue Number261221
Device Lot NumberJ7046C
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/10/2023
Supplement Date Received by Manufacturer04/20/2023
Initial Report FDA Received Date02/28/2023
Supplement Report FDA Received Date05/04/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/23/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age60 YR
Patient SexFemale
-
-