| Model Number |
CAT02590 |
| Health Effect - Clinical Codes |
Foreign Body In Patient (2687); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/30/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the jaw of the device broke during procedure.All pieces were retrieved.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
| |
|
Event or Problem Description
|
|
It was reported that the jaw of the device broke during procedure.All pieces were retrieved.
|
| |
|
Additional Manufacturer Narrative
|
|
This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: jaw broken during surgery probable root cause: design - inadequate raw material selection - tip geometry not designed to facilitate soft tissue penetration manufacturing - instrument tip or needle shaft not manufactured to specification - incorrect raw materials used for manufacture application - user technique related factors - difficult anatomy, extremely tough soft tissue the reported failure mode will be monitored for future reoccurrence.Manufacture date is not known.
|
| |
|