• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE AIR+ MENISCAL SYSTEM - CURVED DOWN; SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE AIR+ MENISCAL SYSTEM - CURVED DOWN; SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE Back to Search Results
Model Number 4723
Health Effect - Clinical Codes Foreign Body In Patient (2687); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/03/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
 
Event or Problem Description
It was reported that the anchor potentially broke and pieces were potentially unable to be retrieved.
 
Event or Problem Description
It was reported that the anchor potentially broke and pieces were potentially unable to be retrieved.
 
Additional Manufacturer Narrative
This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: broken during surgery probable root cause: design - inadequate raw material specification (needle or depth indicator) - needle tube thickness too thin or too thick - suture does not have adequate working length - inadequate handle assembly design - inadequate raw material - depth indicator thickness too thin or too thick process - depth indicator not manufactured to specification - needle or implant body not manufactured to specification - dimple process on needle not manufactured to specification - inadequate design of needle/depth indicator to create interference to retain/remove depth indicator - stiffener not manufactured to specification application - user intentionally bends/forces instrument (g8) - incision too small (g9) - inadequate or improper visualization (g10) - user not familiar with device use (g11) - user does not use sled or other technique to prevent catching on soft tissue (g12) the reported failure mode will be monitored for future reoccurrence.Manufacture date is not known h3 other text : 81.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIR+ MENISCAL SYSTEM - CURVED DOWN
Common Device Name
SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
lucas wolski
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key16455945
Report Number0002936485-2023-00173
Device Sequence Number6452345
Product Code GAT
UDI-Device Identifier07613327482072
UDI-Public07613327482072
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K153087
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/04/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number4723
Device Catalogue Number4723
Device Lot Number7010169
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/03/2023
Supplement Date Received by Manufacturer02/03/2023
Initial Report FDA Received Date02/28/2023
Supplement Report FDA Received Date07/05/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-