• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: IRIDEX CORPORATION MICROPULSE P3; LASER INSTRUMENT, SURGICAL, POWERED, PRODUCT CODE: HQF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

IRIDEX CORPORATION MICROPULSE P3; LASER INSTRUMENT, SURGICAL, POWERED, PRODUCT CODE: HQF Back to Search Results
Model Number 15522
Medical Device Problem Code Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/21/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Iridex became aware of a complaint reporting a patient experiencing a sizzle followed by a smell during treatment with the micropulse p3 probe.A specific root cause could not be determined.The doctor followed the instructions on the micropulse ifu and immediately stopped treatment with the probe.There was no patient injury reported.The treatment was performed within specifications.Based on similar reported events, the most likely cause of the issue is suspected to be contamination (e.G.Blood, tissue or betadine) on the probe tip when the laser was activated or laser activation when the probe is over pigmented conjunctival tissue.Engineering failure analysis was not performed because the device was not returned for evaluation.The micropulse p3 ifu has clear instructions to examine the probe tip before treatment and to avoid areas of perilimbal pigmentation.Per the ifu 'excessive treatment power may result in ocular surface burns or ciliary body hemorrhage.
 
Event or Problem Description
Part a is not filled out as patient identifier details were not provide to iridex.This report has been created to make corrections to the previous report.The changes include: 1) report c: is this a combination product - no 2) report d: updates to section d8, d9 3) part h: update to include that this a correction report 4) part d: update the lot number information.The lot number was filled out in the section for serial number.Updated this section with the udi number.5) part e: updates to the initial reporter section with infromation regarding the complainant 6) part h4: updates to date of manufacture.This information was not submitted previously the following text were provided in the initial report: iridex became aware of a complaint reporting a patient experiencing a sizzle followed by a smell during treatment with the micropulse p3 probe.A specific root cause could not be determined.The doctor followed the instructions on the micropulse ifu and immediately stopped treatment with the probe.There was no patient injury reported.The treatment was performed within specifications.Based on similar reported events, the most likely cause of the issue is suspected to be contamination (e.G.Blood, tissue or betadine) on the probe tip when the laser was activated or laser activation when the probe is over pigmented conjunctival tissue.Engineering failure analysis was not performed because the device was not returned for evaluation.The micropulse p3 ifu has clear instructions to examine the probe tip before treatment and to avoid areas of perilimbal pigmentation.Per the ifu 'excessive treatment power may result in ocular surface burns or ciliary body hemorrhage.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MICROPULSE P3
Common Device Name
LASER INSTRUMENT, SURGICAL, POWERED, PRODUCT CODE: HQF
Manufacturer (Section D)
IRIDEX CORPORATION
1212 terra bella ave.
mountain view CA 94043 1824
Manufacturer (Section G)
IRIDEX CORPORATION
1212 terra bella ave
mountain view CA 94043 9149
Manufacturer Contact
madhumita srikanth
1212 terra bella ave.
mountain view, CA 94043-1824
6502189149
MDR Report Key16456362
Report Number2939653-2022-00036
Device Sequence Number6316226
Product Code HQF
UDI-Device Identifier00813125015251
UDI-Public813125015251
Combination Product (Y/N)Y
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K143154
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number15522
Device Lot Number800760
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date02/28/2023
Supplement Report FDA Received Date10/12/2023
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/05/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-