Additional information was added to d4 (expiration date), h3, h6 and h10: h10: the actual device was not available; however, a photograph of the sample was provided for evaluation.A visual inspection was performed to the photograph using the naked eye which revealed that the inlet tubing detached from the spike body.The reported condition was verified.The cause of the condition could not be determined; however, the most likely causes are due to inadequate or lack of cyclohexanone applied at the spike/tubing connection in the manufacturing process or due to customer handing/usage.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|