| Model Number |
INNOVANCE D-DIMER |
| Medical Device Problem Code |
Use of Device Problem (1670)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/01/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A customer reported that a physician misinterpreted an innovance d-dimer result of 0.93 mcg/ml with a reference range of 0.19 to 1.12 mcg/ml from the sysmex cs-2500 system as a result within the normal range, instead of above the cut-off for the exclusion of venous thromboembolism (vte).As a result, the physician(s) sent the patient home without further testing.After realizing the misinterpretation, the physician(s) re-evaluated the case and determined that the patient was not at risk of vte.This report is being filed in an abundance of caution.There are no known reports of patient intervention or adverse health consequences due to this event.
|
| |
|
Additional Manufacturer Narrative
|
|
A united states (us) customer contacted a siemens customer care center (ccc).Quality controls (qc) recovered in range at the time of the event.The customer informed siemens that they updated their reference range to be 0.19 to - 0.49 mg/l to utilize the cut-off when assessing for venous thromboembolism (vte).Siemens evaluated the innovance d-dimer instructions for use (ifu) and sysmex automated blood coagulation analyzer cs-2500 reference guide and determined that the labeling regarding the cut-off value for the exclusion of vte using the innovance d-dimer is not misleading.The following sections demonstrates the assay was evaluated to validate the exclusion of dvt: "clinical performance of the innovance d-dimer assay to exclude dvt" in the innovance d-dimer ifu and "clinical performance" in the sysmex automated blood coagulation analyzer cs-2500 reference guide.Both document indicates patient specimens were tested with the innovance d-dimer assay and results were compared to a cut-off value of 0.50 mg/l (feu).A d-dimer result < 0.50 mg/l (feu) was considered negative and a d-dimer result >= 0.50 mg/l (feu) was considered positive.The cause of the event is use error and this report is being filed in an abundance of caution as there is no evidence of an instrument nor reagent malfunction.The reagent is performing according to specifications.No further evaluation of this device is required.
|
| |
|
Search Alerts/Recalls
|