| Model Number |
X SERIES |
| Medical Device Problem Code |
No Display/Image (1183)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during a routine shift check by a clinician, the device's display was distorted and no clinical information could be seen.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to zoll medical corporation for evaluation.During disassembly of the device to determine root cause, the technician observed damage not consistent with typical use.A visual inspection of the device found extensive damage to the lcd display and front enclosure not consistent with normal wear and tear.The lcd display and front enclosure were replaced.The device was recertified and returned to the customer.Analysis of reports of this type has not identified an increase in trend.
|
| |
|