TIDI PRODUCTS LLC TELEFLEX GRIP-LOK SPECIALTY SECUREMENT DEVICE, CVC & PICC; DEVICE, INTRAVASCULAR CATHETER SECUREMENT Back to Search Results
Model Number 3604MCS-TA-1
Medical Device Problem Code Loss of or Failure to Bond (1068)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The complaint product was not available for evaluation.Retain samples from 3 of the last 4 lots shipped to this customer were evaluated.Multiple retain samples were tested for each of the 3 lots pulled for adhesive functionality via hanging weight test.All reviewed samples passed the test and functioned per original release specifications.Additionally, the samples were applied to the tester¿s skin to see if they would stick to the skin and all did.The reported complaint could not be duplicated.The reporter was contacted for more information regarding the event and the users and it was reported that they are new users who are using the grip-loks because of a shortage of the previously used device: piccline statlock ref: (b)(4) bd.Similar complaints have been reported from former stat-lok users due to the aggressive adhesive used in that device.The reporter (distributor) was asked if any training had been provided to the end user or if training is planned, but no response was received.At this time, it appears that the root cause is related to new users and device training problem.The reporter has been provided with a copy of the current instructions for use to be forwarded to the customer as necessary.No further corrective or preventive actions are required currently.All complaints are trended and reviewed by management on a monthly basis.As part of this monthly review, any excursion above the control limits for this failure mode will be assessed, documented and acted upon as warranted.Manufacturer reference file (b)(4).
 
Event or Problem Description
It was reported by teleflex: not holding / device does not stick to the skin: "this complaint covers the complaint with no details.Dressing with ineffective attachment to the patient's skin.All the lot numbers are involved.¿.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TELEFLEX GRIP-LOK SPECIALTY SECUREMENT DEVICE, CVC & PICC
Common Device Name
DEVICE, INTRAVASCULAR CATHETER SECUREMENT
Manufacturer (Section D)
TIDI PRODUCTS LLC
570 enterprise drive
neenah WI 54956
Manufacturer Contact
chris rahn
570 enterprise drive
neenah, WI 54956
9207514300
MDR Report Key16463807
Report Number2182318-2023-00011
Device Sequence Number11386142
Product Code KMK
UDI-Device Identifier00618125160945
UDI-Public00618125160945
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 02/01/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number3604MCS-TA-1
Device Catalogue Number3604MCS-TA-1
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/01/2023
Initial Report FDA Received Date03/01/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1