| Model Number |
ASDS-14F-075 |
| Medical Device Problem Code |
Air/Gas in Device (4062)
|
| Health Effect - Clinical Code |
Non specific EKG/ECG Changes (1817)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Event or Problem Description
|
|
It was reported that on an unknown date, an unknown device was chosen for implant with an amplatzer steerable delivery sheath.It was reported the patient exhibited minor st elevations due to air that resolved without intervention.No additional information was provided.
|
| |
|
Event or Problem Description
|
|
This is being filed as a physician reported that there were about 6 patients who had st elevation that resolved within a few minutes and no interventions.It was suspected that the st elevation was due to air but was not confirmed as there was no confirmation of air visualized in the patient or the delivery system. it was reported that on an unknown date, an unknown device was chosen for implant with an amplatzer steerable delivery sheath.It was reported the patient exhibited minor st elevations that resolved quickly without intervention.Air was able to be visualized on fluoroscopy imaging.It was noted that there were no major adverse patient effects.
|
| |
|
Additional Manufacturer Narrative
|
|
An event of minor st elevations was reported.A returned device inspection to rule out any device-related causes could not be performed as the device was not returned for analysis.Information from the field indicated that the reported event resolved itself quickly.The field also indicated that the physician was able to visualize what looked like air on the fluoro images.No additional information was available for review.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
| |
|
Search Alerts/Recalls
|