• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SISTEMAS MEDICOS ALARIS, S.A. DE C.V. BD MAXGUARD¿ FLOW CONTROLLER ADMINISTRATION SET WITH NEEDLELESS Y-SITE(S); INTRAVASCULAR ADMINISTRATION SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SISTEMAS MEDICOS ALARIS, S.A. DE C.V. BD MAXGUARD¿ FLOW CONTROLLER ADMINISTRATION SET WITH NEEDLELESS Y-SITE(S); INTRAVASCULAR ADMINISTRATION SET Back to Search Results
Model Number MFS102
Medical Device Problem Code Complete Blockage (1094)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/27/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported that the bd maxguard¿ flow controller administration set with needleless y-site(s) was occluded during use.The following information was provided by the initial reporter: " situation: fluid bag is spiked; the flow controller is set to open and the line primes without issue.When the controller is set to the rate decide there is no flow.Occurred 7 times in last tow days.Rns moved to another device or used alaris pump.Background: the device is designed to be used in open mode to prime the tubing and then the rn can set the rate of flow per ordered rate.".
 
Event or Problem Description
It was reported that the bd maxguard¿ flow controller administration set with needleless y-site(s) was occluded during use.The following information was provided by the initial reporter: " situation: fluid bag is spiked; the flow controller is set to open and the line primes without issue.When the controller is set to the rate decide there is no flow.Occurred 7 times in last tow days.Rns moved to another device or used alaris pump.Background: the device is designed to be used in open mode to prime the tubing and then the rn can set the rate of flow per ordered rate.".
 
Additional Manufacturer Narrative
H6: investigation summary no sample ( mat # mfs102) received by the customer.Photos representation was provided for the complaint.It was reported by the customer that the product is having flow issues.However the failure could not be verified in the photos.A device history record review for model mfs102 and lot number 22029883 was performed.There were no quality notifications issued for the failure mode reported by the customer during the production build of this lot.Due to no sample being received a root cause could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD MAXGUARD¿ FLOW CONTROLLER ADMINISTRATION SET WITH NEEDLELESS Y-SITE(S)
Common Device Name
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer (Section G)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8448235433
MDR Report Key16464669
Report Number9616066-2023-00288
Device Sequence Number7482181
Product Code FPA
UDI-Device Identifier10885403236334
UDI-Public10885403236334
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K051499
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2025
Device Model NumberMFS102
Device Catalogue NumberMFS102
Device Lot Number22029883
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/07/2023
Supplement Date Received by Manufacturer04/05/2023
Initial Report FDA Received Date03/01/2023
Supplement Report FDA Received Date04/11/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/28/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1