| Model Number |
275510 |
| Medical Device Problem Code |
Failure to Cut (2587)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Several attempts were made to obtain the device for inspection and investigation, but it was not returned to the manufacturer.The manufacturing lot number is unknown and it is impossible to review device manufacturing history.However, it has been established that the device was manufactured in 2016.Possibly the problem was due to normal wear over time.The complaint is considered closed.Should additional information become available, it will be reopened.
|
| |
|
Event or Problem Description
|
|
The wire cutter was used to cut k-wire, but the ends of the k-wire frayed and small fragments broke off into the patient during surgery.The broken pieces were recovered.No patient injury occurred; however, there was an unspecified delay in surgery time.
|
| |
|