DEPUY SYNTHES PRODUCTS LLC 14 CM QD ANGLE ATTACHMENT; MOTOR, DRILL, ELECTRIC Back to Search Results
Model Number QD14-G1
Medical Device Problem Codes Complete Loss of Power (4015); Excessive Heating (4030)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
It was reported by switzerland that during service and evaluation, it was determined that the attachment device produced heat.It was further determined that the device failed pretest for temperature assessment.It was noted in the service order that during pre-surgery, it was discovered that the device did not work anymore.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.All available information has been disclosed.If additional information should available, a supplemental medwatch will be submitted accordingly.
 
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.Device evaluation: the actual device was returned for evaluation.During repair, it was determined that the reported condition that the device did not work anymore was not confirmed.Therefore, an assignable root cause for the reported condition of does not function was not determined.However, during evaluation, it was determined that the device produced heat.The assignable root cause was determined to be traced to user, which is user error.Udi: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
14 CM QD ANGLE ATTACHMENT
Common Device Name
MOTOR, DRILL, ELECTRIC
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
4500 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
kate karberg
4500 riverside drive
palm beach gardens, FL 
3035526892
MDR Report Key16465974
Report Number1045834-2023-00295
Device Sequence Number20777055
Product Code HBC
UDI-Device Identifier00845384016458
UDI-Public00845384016458
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
K133604
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 02/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberQD14-G1
Device Catalogue NumberQD14-G1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/15/2023
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/27/2023
Initial Report FDA Received Date03/02/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/19/2016
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-