| Model Number |
1010 |
| Medical Device Problem Code |
Interrogation Problem (4017)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/27/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported the device could not be interrogated.Technical services had the caller select "scan for device" on the programmer, and the interrogation was successful.The device was then explanted and replaced for normal battery depletion.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon receipt at our post market quality assurance laboratory, the device was examined.Visual inspection noted no damage.The device was then exposed to simulated heart load conditions, and the defibrillation functions were tested.The device operated appropriately, according to its performance specifications with no interruptions in therapy output or programmer communication at the returned programmed settings.A series of electrical tests was also performed, and again, normal device function was observed.Laboratory analysis did not identify any characteristics that would have caused or contributed to the reported clinical observations.
|
| |
|
Event or Problem Description
|
|
This supplemental report is being filed to provide additional information regarding product analysis.It was reported the device could not be interrogated.Technical services had the caller select "scan for device" on the programmer, and the interrogation was successful.The device was then explanted and replaced for normal battery depletion.
|
| |
|