MAUDE Adverse Event Report: ARROW INTERNATIONAL LLC REDIGUARD IAB: 8FR 40CC; SYSTEM, BALLOON, INTRA-AORTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL LLC REDIGUARD IAB: 8FR 40CC; SYSTEM, BALLOON, INTRA-AORTIC Back to Search Results
Model Number IPN038077
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/20/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that, "highly torturous femoral artery, the sheath and the shaft of iabp was kinked and the balloon was not inflating, balloon was removed as bp was not displayed".Additional information states that a 2nd iab was used and inserted at the same insertion site.No harm or injury to the patient.The patient status is reported as "fine".
 
Additional Manufacturer Narrative
Qn# (b)(4).N/a.Other remarks: n/a.Corrected data: n/a.
 
Additional Manufacturer Narrative
(b)(4).Complaint verification testing could not be performed as no sample was returned for analysis.A device history record review was performed, and no relevant findings were identified.Without the device to evaluate, the complaint could not be confirmed, and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.Other remarks: n/a.Corrected data: n/a.
 
Event or Problem Description
It was reported that, "highly torturous femoral artery, the sheath and the shaft of iabp was kinked and the balloon was not inflating, balloon was removed as bp was not displayed".Additional information states that a 2nd iab was used and inserted at the same insertion site.No harm or injury to the patient.The patient status is reported as "fine".
 
Additional Manufacturer Narrative
(b)(4).The lot number reported is 18f22j0014.Returned for investigation was a 40cc 8fr rediguard in-tra-aortic balloon catheter (iabc) without the original packaging.The sample was returned in a cardboard box and was in a sealed ziploc bag.Upon return, the peel away sheath extrusion was noted cut.The one-way valve was tethered to the short driveline tubing.The bladder was fully unwrapped.More than 15 kinks to the iabc central lumen were noted throughout the length of the iabc.No blood was noted on the interior or the exterior surfaces of the returned sample.The sample was likely cleaned prior to its return.No other visual damage or abnormalities were not-ed to the returned sample.The one-way valve was tested and passed.A vacuum was pulled on the one-way valve, and it held for at least 1 minute according to quality system document.The iabc central lumen was aspirated and flushed using a 60cc lab-inventory s yringe.No abnormali-ties or debris were noted.The iabc was leak tested using in accordance with testing methods from manufacturing procedure.No leaks were detected.Full inflation was achieved.The device passed the leak test.A lab inventory 0.025in guidewire was back loaded through the iabc distal tip.Resistance was noted at the locations of the previously noted kinks.The guidewire was able to advance through the central lumen.No blood or debris was noted.A device history record (dhr) review was conducted for the lot number with no relevant findings.The device passed all manufacturing specifications prior to release.The reported complaint that the "shaft of iabp was kinked" is confirmed.The intra-aortic balloon catheter (iabc) was returned with multiple kinks to the central lumen.The iabc bladder was fully intact with no damage noted.Based on a review of the device history record (dhr), the product met specification upon release; however, the specifications were not met during the complaint investigation due to the kinked central lumen.The root cause of the kinked central lumen is undetermined; a potential root cause is customer handling related.No further action required at this time.This will be monitored for any developing trends.Other remarks: n/a.Corrected data: n/a.
 
Event or Problem Description
It was reported that, "highly torturous femoral artery, the sheath and the shaft of iabp was kinked and the balloon was not inflating, balloon was removed as bp was not displayed".Additional information states that a 2nd iab was used and inserted at the same insertion site.No harm or injury to the patient.The patient status is reported as "fine".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REDIGUARD IAB: 8FR 40CC
Common Device Name
SYSTEM, BALLOON, INTRA-AORTIC
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL INC.
16 elizabeth drive
chelmsford MA 01824
Manufacturer Contact
sachin sony
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key16467192
Report Number3010532612-2023-00130
Device Sequence Number7482263
Product Code DSP
UDI-Device Identifier10801902172942
UDI-Public(01)10801902172942(11)220913(17)240831(10)18F22J0014(21)
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K981660
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2024
Device Model NumberIPN038077
Device Catalogue NumberIAB-S840C
Device Lot Number18F22J0014
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/20/2023
Supplement Date Received by Manufacturer04/11/2023
02/05/2026
Initial Report FDA Received Date03/02/2023
Supplement Report FDA Received Date04/12/2023
02/06/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/13/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexFemale
-
-