• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HOUSTON CONSTELLATION VISION SYSTEM, ACCESSORY, IRRIGATION/ASPIRATION TUBING SET; UNIT, PHACOFRAGMENTATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HOUSTON CONSTELLATION VISION SYSTEM, ACCESSORY, IRRIGATION/ASPIRATION TUBING SET; UNIT, PHACOFRAGMENTATION Back to Search Results
Catalog Number 8065750918
Medical Device Problem Code Disconnection (1171)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/30/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A nurse reported that the irrigation aspiration (ia) tube was not tightly connected with the handpiece and was easily falling off during the phacoemulsification (phaco) phase of the cataract surgery.The tube was replaced with a new tube which worked normally, and the surgery was completed.There was no patient harm.
 
Additional Manufacturer Narrative
A review of the device history record traceable to the reported lot number indicates that the product was processed and released according to the product¿s acceptance criteria.The returned sample was visually inspected, and no obvious defects were found.The aspiration and irrigation fittings could be connected into the handpiece.The irrigation aspiration (i/a) manifold was tested with the lab stock cassette and other components.The i/a manifold passed priming and tuning process.The aspiration and irrigation fittings seated in the handpiece securely and firmly during testing.The root cause of the customer's complaint could not be established; the returned sample met specifications.After an investigation of this complaint, it has been determined that this sample functioned per specifications; therefore, no corrective action is required at this time.No adverse trends have been observed associated with the reported product and event.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION VISION SYSTEM, ACCESSORY, IRRIGATION/ASPIRATION TUBING SET
Common Device Name
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer (Section G)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key16467587
Report Number1644019-2023-00208
Device Sequence Number17438757
Product Code HQC
UDI-Device Identifier00380657509188
UDI-Public00380657509188
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K101285
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 05/31/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2023
Device Catalogue Number8065750918
Device Lot Number2454619H
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/02/2023
Supplement Date Received by Manufacturer05/05/2023
Initial Report FDA Received Date03/02/2023
Supplement Report FDA Received Date05/31/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/03/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-