COLLAGEN MATRIX, INC GIBSON HEALTHCARE BONE GRAFT PLUG; SYNOSS COLLAGEN SYNTHETIC MATERIAL - COLLAGEN BONE GRAFT MATRIX Back to Search Results
Catalog Number 5824-413
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Pain (1994)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional quality control testing of samples is pending.
 
Event or Problem Description
Customer initially reported the gibson bone graft plug was causing healing issues with "a number of patients." additional, limited information was provided.Customer stated at least five patients experienced healing issues.For patient 1 of 5, the bone graft appeared to quickly swell, appear infected and extrude from the extraction site.A few days after implantation (implant date not provided), patient returned to the office and reported having a "terrible taste and smell." medical intervention was required to remove the graft and clean the extraction site.After the initial intervention, customer reported that some patients had to return to the office "a few times" to remove the graft material that continued to extrude and clean and irrigate the sites.It was not specified how many patients had to return to the office, nor was it specified how many times the patients had to return to the office.Customer stated this caused delayed healing and patient pain.Customer confirmed "patients are fine now." however, there was no information provided on current patient status, other specific patient information, and outcome of event.No further information was provided.
 
Additional Manufacturer Narrative
Additional quality control testing showed the product met acceptable sterility criteria.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GIBSON HEALTHCARE BONE GRAFT PLUG
Common Device Name
SYNOSS COLLAGEN SYNTHETIC MATERIAL - COLLAGEN BONE GRAFT MATRIX
Manufacturer (Section D)
COLLAGEN MATRIX, INC
15 thornton rd
oakland NJ 07436
Manufacturer (Section G)
COLLAGEN MATRIX, INC
15 thornton rd
oakland NJ 07436
Manufacturer Contact
peggy hansen
15 thornton road
oakland, NJ 07436
MDR Report Key16467630
Report Number2249852-2023-00004
Device Sequence Number5263533
Product Code LYC
UDI-Device Identifier00813954024233
UDI-Public(01)00813954024233(17)240731(10)SMCPN210601
Combination Product (Y/N)N
Initial Reporter StateUT
Initial Reporter CountryUS
PMA/510(K) Number
K083742
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date (Section B) 06/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Expiration Date07/31/2024
Device Catalogue Number5824-413
Device Lot NumberSMCPN210601
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/01/2023
Supplement Date Received by Manufacturer06/02/2023
Initial Report FDA Received Date03/02/2023
Supplement Report FDA Received Date06/13/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/31/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-