MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ELUVIA DRUG-ELUTING VASCULAR STENT SYSTEM; STENT, SUPERFICIAL FEMORAL ARTERY, DRUG-ELUTING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ELUVIA DRUG-ELUTING VASCULAR STENT SYSTEM; STENT, SUPERFICIAL FEMORAL ARTERY, DRUG-ELUTING Back to Search Results
Model Number 24657
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Vascular Dissection (3160); Thrombosis/Thrombus (4440)
Date of Event 02/03/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Patient identifier: (b)(6).Age at time of event: the patient was 59 years old at the time of study enrollment.
 
Event or Problem Description
S2444 elegance clinical study.It was reported that thrombosis and dissection occurred during the index procedure.On (b)(6) 2023, the subject underwent treatment with eluvia drug-eluting stent and ranger drug coated balloons as a part of the elegance clinical trial.The target lesion #001 was in the right proximal superficial femoral artery 9sfa) extending into right mid sfa with proximal reference vessel diameter of 6 mm and distal reference vessel diameter of 7 mm with lesion length of 40 mm and 70% stenosis and was classified as tasc ii a lesion.Treatment of target lesion was performed by study devices, 7 mm x 60 mm eluvia drug-eluting stent and 6 mm x 40 mm ranger drug-coated balloon.Following post dilation with a non-bsc balloon was performed, the final residual stenosis was noted to be 30%.The target lesion #002 was in the right common femoral artery with proximal reference vessel diameter of 7 mm and distal reference vessel diameter of 7 mm with lesion length of 60 mm and 80% stenosis and was classified as tasc ii b lesion.Treatment of target lesion was performed by study device, 7 mm x 60 mm ranger drug-coated balloon.Following treatment, the final residual stenosis was noted to be 30%.On the same day of index procedure, during the treatment of target lesions right proximal sfa and right mid sfa, dissection of grade a and thrombosis were noted and in right common femoral artery.Per target lesions 001 and 002, balloon dilation and bailout stent were performed to treat dissection and thrombosis noted.In addition, medication was given, angiography and thrombectomy were also performed.No further patient consequences were reported.
 
Additional Manufacturer Narrative
A1 - patient identifier: (b)(6).A2 - age at time of event: the patient was 59 years old at the time of study enrollment.Updated fields: b5 - additional information was received that confirmed the originally reported event of thrombosis was not related to the complaint device.Thus, the event summary has been updated accordingly.G1 - mfr contact first name, mfr contact last name, mfr contact phone number, mfr contact email.H6 - removal of thrombosis/thrombus e0514; updated evaluation method codes, evaluation result codes, and evaluation conclusion codes.
 
Event or Problem Description
S2444 elegance clinical study.It was reported that thrombosis and dissection occurred during the index procedure.On (b)(6) 2023, the subject underwent treatment with eluvia drug-eluting stent and ranger drug coated balloons as a part of the elegance clinical trial.The target lesion #001 was in the right proximal superficial femoral artery 9sfa) extending into right mid sfa with proximal reference vessel diameter of 6 mm and distal reference vessel diameter of 7 mm with lesion length of 40 mm and 70% stenosis and was classified as tasc ii a lesion.Treatment of target lesion was performed by study devices, 7 mm x 60 mm eluvia drug-eluting stent and 6 mm x 40 mm ranger drug-coated balloon.Following post dilation with a non-bsc balloon was performed, the final residual stenosis was noted to be 30%.The target lesion #002 was in the right common femoral artery with proximal reference vessel diameter of 7 mm and distal reference vessel diameter of 7 mm with lesion length of 60 mm and 80% stenosis and was classified as tasc ii b lesion.Treatment of target lesion was performed by study device, 7 mm x 60 mm ranger drug-coated balloon.Following treatment, the final residual stenosis was noted to be 30%.On the same day of index procedure, during the treatment of target lesions right proximal sfa and right mid sfa, dissection of grade a and thrombosis were noted and in right common femoral artery.Per target lesions 001 and 002, balloon dilation and bailout stent were performed to treat dissection and thrombosis noted.In addition, medication was given, angiography and thrombectomy were also performed.No further patient consequences were reported.It was further reported that the thrombosis that occurred on (b)(6) 2023, was not related to the study device or study procedure.
 
Event or Problem Description
(b)(4) clinical study.It was reported that thrombosis and dissection occurred during the index procedure.On (b)(6) 2023, the subject underwent treatment with eluvia drug-eluting stent and ranger drug coated balloons as a part of the elegance clinical trial.The target lesion #(b)(4) was in the right proximal superficial femoral artery 9sfa) extending into right mid sfa with proximal reference vessel diameter of 6 mm and distal reference vessel diameter of 7 mm with lesion length of 40 mm and 70% stenosis and was classified as tasc ii a lesion.Treatment of target lesion was performed by study devices, 7 mm x 60 mm eluvia drug-eluting stent and 6 mm x 40 mm ranger drug-coated balloon.Following post dilation with a non-bsc balloon was performed, the final residual stenosis was noted to be 30%.The target lesion #(b)(4) was in the right common femoral artery with proximal reference vessel diameter of 7 mm and distal reference vessel diameter of 7 mm with lesion length of 60 mm and 80% stenosis and was classified as tasc ii b lesion.Treatment of target lesion was performed by study device, 7 mm x 60 mm ranger drug-coated balloon.Following treatment, the final residual stenosis was noted to be 30%.On the same day of index procedure, during the treatment of target lesions right proximal sfa and right mid sfa, dissection of grade a and thrombosis were noted and in right common femoral artery.Per target lesions (b)(4), balloon dilation and bailout stent were performed to treat dissection and thrombosis noted.In addition, medication was given, angiography and thrombectomy were also performed.No further patient consequences were reported.It was further reported that the thrombosis that occurred on (b)(6) 2023, was not related to the study device or study procedure.It was further reported that during the index procedure on (b)(6) 2023, target lesion (b)(4) was treated with 6 mm x 60 mm ranger drug-coated balloon.The target lesion (b)(4) was classified as tasc ii a lesion.During the index procedure, dissection of grade e was noted in the right cfa due to user error and in response, balloon dilation was performed using 7 mm x 60 mm ranger dcb followed by placement of non-boston scientific stent.Post procedure, the final residual stenosis was noted to be 30%.On (b)(6) 2023, the subject was discharged on dual anti-platelet therapy.
 
Additional Manufacturer Narrative
A1 - patient identifier: (b)(6).A2 - age at time of event: the patient was 59 years old at the time of study enrollment.Updated fields: b5 - additional information was received that confirmed the originally reported event of thrombosis was not related to the complaint device.Thus, the event summary has been updated accordingly.G1 - mfr contact first name, mfr contact last name, mfr contact phone number, mfr contact email h6 - removal of thrombosis/thrombus e0514; updated evaluation method codes, evaluation result codes, and evaluation conclusion codes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELUVIA DRUG-ELUTING VASCULAR STENT SYSTEM
Common Device Name
STENT, SUPERFICIAL FEMORAL ARTERY, DRUG-ELUTING
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key16468653
Report Number2124215-2023-09313
Device Sequence Number20129312
Product Code NIU
UDI-Device Identifier08714729876694
UDI-Public08714729876694
Combination Product (Y/N)Y
Initial Reporter StateRI
Initial Reporter CountryUS
PMA/510(K) Number
P180011
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Study,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/04/2024
Device Model Number24657
Device Catalogue Number24657
Device Lot Number0029410109
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/08/2023
Supplement Date Received by Manufacturer03/09/2023
07/18/2024
Initial Report FDA Received Date03/02/2023
Supplement Report FDA Received Date03/31/2023
08/08/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/05/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
RANGER PACLITAXEL COATED PTA BALLOON CATHETER; RANGER PACLITAXEL COATED PTA BALLOON CATHETER; RANGER PACLITAXEL COATED PTA BALLOON CATHETER; RANGER PACLITAXEL COATED PTA BALLOON CATHETER; RANGER PACLITAXEL COATED PTA BALLOON CATHETER; RANGER PACLITAXEL COATED PTA BALLOON CATHETER
Outcome Attributed to Adverse Event Other;
Patient SexFemale
Patient RaceWhite
-
-