| Model Number |
8900-0223-01 |
| Medical Device Problem Code |
Unable to Obtain Readings (1516)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during a routine shift check by a clinician, the associated device displayed a "poor pad contact" message using these electrode pads.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The electrodes were returned to zoll medical corporation for evaluation.The customer's report was verified and attributed to a disconnected self-test wire connecting the two electrodes.This type of failure does not disable the electrode from delivering therapy when attached to a defibrillator.This is a malfunction which only impacts the self-test of the electrode with the defibrillator.The electrodes were scrapped and replaced.Analysis of reports of this type has not identified an increase in trend.
|
| |
|