ALLERGAN (COSTA RICA) STYLE 68 SALINE FILLED BREAST IMPLANT; PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE Back to Search Results
Model Number 68LP-300
Medical Device Problem Code Device Appears to Trigger Rejection (1524)
Health Effect - Clinical Code Capsular Contracture (1761)
Date of Event 02/08/2023
Type of Reportable Event Serious Injury
Event or Problem Description
Healthcare professional reported a right side capsular contracture baker grade iii/iv.Device has been explanted.
 
Additional Manufacturer Narrative
A review of the device history record has been completed.No deviations or non-conformances noted.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture baker grade iii/iv.
 
Additional Manufacturer Narrative
Device evaluation: based on the device analysis grid, the assessments of the complaint are: ¿ capsular contracture: unable to observe as it is a medical event not related to the device.Additional observations: ¿ white particles observed on the device.¿ observed opening and broken on anterior side assessed as surgical damage.¿ missing shell assessed as inconclusive.No further actions are required as the device was implanted.
 
Event or Problem Description
Healthcare professional reported a right side capsular contracture baker grade iii/iv.Device has been explanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STYLE 68 SALINE FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer (Section G)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer Contact
terry ingram
12331-a riata trace parkway
building 3
austin, TX 78727
8479366324
MDR Report Key16471392
Report Number9617229-2023-02782
Device Sequence Number15397274
Product Code FWM
UDI-Device Identifier10888628000582
UDI-Public10888628000582
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P990074
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2023
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/23/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/03/2025
Device Model Number68LP-300
Device Catalogue Number68LP-300
Device Lot Number3573705
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/09/2023
Supplement Date Received by Manufacturer06/01/2023
Initial Report FDA Received Date03/02/2023
Supplement Report FDA Received Date06/23/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/03/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age38 YR
Patient SexFemale
Patient Weight68 KG
-
-