| Model Number |
670170 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/07/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when required.
|
| |
|
Event or Problem Description
|
|
It was reported that a cuff leak was discovered while in use with a patient.Leakage confirmed when product was withdrawn from patient and further tested.Customer states there was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by manufacturer and h6.Health impact, and evaluation codes: updated.One device was received for investigation.The reported issue was confirmed during functional testing, which detected a leak in the sample device cuff.A review of the manufacturing device history records found no non-conformances, however the investigation attributed this observed condition to a manufacturing issue.This issue will continue to be monitored and further actions taken accordingly.
|
| |
|