| Model Number |
8900-0800-01 |
| Medical Device Problem Code |
Unable to Obtain Readings (1516)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during biomed testing, the associated defibrillator was unable to detect these attached electrode pads.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The electrodes were returned to zoll taiwan for evaluation.The customer's report was observed and attributed to faulty electrodes.The electrodes were replaced to resolve the report.Analysis of reports of this type has not identified an increase in trend.
|
| |
|