| Model Number |
S-55-060-120-P6 |
| Medical Device Problem Codes |
Break (1069); Difficult to Remove (1528); Stretched (1601); Migration (4003)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
02/10/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents from this lot.The investigation determined the reported difficulties appear to be related to circumstances of the procedure as it is likely that during removal interaction between the deployed stent and the device tip resulted in the reported difficult to remove.Manipulation to remove the device resulted in the reported stretched stent and the reported stent migration.Inadvertent mishandling possibly resulted in the reported handle break; however this cannot be confirmed.The investigation was unable to determine a conclusive cause for the reported break.There is no indication of a product quality issue with respect to manufacture, design or labeling.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure was performed to treat a lesion in the popliteal artery.A 5.5x60mm supera self-expanding stent was deployed; however, during removal of the self-expanding stent system (sess) resistance between the stent and tip was felt.After several attempts the sess was able to be removed, but the stent elongated and moved approximately 5cm out of the target lesion into the femoral artery.It was further reported that the handle of the system was observed to be damaged with the middle part being open.The procedure was finished and nothing further was done.No additional information was provided.
|
| |
|
Search Alerts/Recalls
|