Investigation: visual inspection: during the investigation, no significant deformations or damage of the valve was determined.Permeability test: a permeability test has shown that the valve is permeable.Computer controlled test: to investigate the claim of under-drainage, the opening pressure is measured using a miethke computer controlled testing apparatus which simulates a cerebrospinal fluid flow.The valve is tested in both the horizontal as well as the vertical positions.The results show that the valve is operating within the accepted tolerance in the horizontal but not in the vertical position.An accelerated outflow of paedigav could be determined.Internal inspection: after dismantling of the valve, deposits were found in paedigav.Results: based on our investigation results, we can determine an accelerated outflow in the valve.The determined deposits can be named as the cause for the accelerated outflow.Proteins in the cerebrospinal fluid can influence the function temporarily and are known side effects in hydrocephalus therapy.We can exclude a defect at the time of release.The valve met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.No further regulatory actions are required from our point of view.
|