|
PERFUSION SYSTEMS BIO-MEDICUS NG 21 FR VENOUS CANNULA; CATHETER, CANNULA AND TUBING, VASCULAR, CARDIO
|
Back to Search Results |
|
| Model Number |
96670-121 |
| Medical Device Problem Code |
Crack (1135)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/07/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Medtronic received information that at an unspecified time it was noted that this bio-medicus nextgen femoral bi-caval venous cannula was cracked at the interface.There was no reported adverse patient effect.Information was requested regarding whether there was a leak from the cannula, if there was blood loss, the amount of blood loss and if a transfusion was required but the information was not provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusion: after investigation at medtronic, the complaint was confirmed for a cracked cannula body connector.Returned product ana lysis confirmed the damage to the connector and the hemostasis cap was not returned with the cannula.It is possible the hemostasis cap was not used when the introducer was placed into the cannula body prior to use and therefore, this complaint would most likely be the result of a use-related issue.This damage to the cannula connector can occur when the introducer is inserted into the cannula body without the use of the hemostasis cap.The introducer handle can cause stretching, crazing or cracking when it is forced into the connector.The issue is more likely to occur with the arterial cannulae models with a clear hemostasis cap that is attached to the introducer on the provided protective sheath, however, the issue can also occur in the venous cannulae models with the blue hemostasis cap.Review of the device history record was not completed as there is no evidence to suggest manufacturing issues with the complaint device.Manufacturing investigation was completed for this failure mode.In this investigation the manufacturing process was reviewed, and the handling of the affected part was minimum, just pick and place into the packaging, no additional assembly.Review of in-process showed that pieces produced are in good condition.Product event notification and awareness was given to the pro duction personnel about the implications and consequences that reported defect has on the field.There were no adverse patient effects as a result of this incidence.Medtronic will continue to monitor for future occurrences and trends per trend analysis procedure.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation summary: visual inspection shows evidence of damage/crack in the connector.Reason for return was confirmed.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|
|
|