• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTEGRA LIFESCIENCES (IRELAND) LTD. DUO HEADLIGHT 2 BAY SYSTEM - US; PFM05

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTEGRA LIFESCIENCES (IRELAND) LTD. DUO HEADLIGHT 2 BAY SYSTEM - US; PFM05 Back to Search Results
Model Number 90620US
Medical Device Problem Code Overheating of Device (1437)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The duo headlight 2 bay system (90620us) was returned for evaluation: failure analysis - the returned duo headlight (90620us) was received is in used condition.Evaluation found that the unit's power cord was losing connection, leading to the fan malfunction and also leading to the light flickering.The power cord will be replaced to address this issue.Root cause - evaluation identified that the reported issue was most likely due to routine use and wear.The reported complaint was confirmed.No manufacturing, workmanship, or material deficiency has been identified.
 
Event or Problem Description
A facility reported that the light of duo headlight 2 bay system (90620us) was flickering on and off.Additionally, the led lamp seemed to be getting excessively hot.No patient injury, death or surgical delay was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DUO HEADLIGHT 2 BAY SYSTEM - US
Common Device Name
PFM05
Manufacturer (Section D)
INTEGRA LIFESCIENCES (IRELAND) LTD.
sragh tullamore, co. offaly
EI 
Manufacturer (Section G)
INTEGRA LIFESCIENCES (IRELAND) LTD.
sragh tullamore, co. offaly
EI  
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key16477784
Report Number3006697299-2023-00021
Device Sequence Number14240552
Product Code FSR
UDI-Device Identifier10381780490746
UDI-Public10381780490746
Combination Product (Y/N)N
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number90620US
Device Catalogue Number90620US
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/26/2023
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/08/2023
Initial Report FDA Received Date03/03/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-