| Model Number |
823184 |
| Medical Device Problem Code |
Obstruction of Flow (2423)
|
| Health Effect - Clinical Code |
Hemorrhage/Blood Loss/Bleeding (1888)
|
| Date of Event |
02/08/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
1 of 2 reports.Other mfg report number: 3013886523-2023-00072.A physician reported a hakim valve (id 823184) and a bactiseal catheter kit (id 823072) were implanted on (b)(6) 2022.A shunt blockage and malfunction occurred on february 07, 2023.The patient had interventricular hemorrhage on the fifth (5th) day after implantation.Due to blockage and malfunction the valve and the catheter were removed and replaced on (b)(6) 2023.
|
| |
|
Additional Manufacturer Narrative
|
|
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
The hakim valve (id 823184) was not returned for evaluation after three good faith attempts (gfes) were made.Lot number information has not been provided; therefore, an evaluation of the device could not be performed, and device history record (dhr) could not be reviewed.The root cause(s) of the reported issue could not be determined.However, a probable root cause for the reported complaint is biological debris and protein build up interfering with the valve.As no valve was returned for investigation this could not be confirmed.If additional relevant information becomes available in the future, this complaint will be reopened, and the respective evaluation performed.Trends will be monitored for this and similar issues.At present, we consider this complaint to be closed.
|
| |
|