• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC, INC. SENSIA IPG; PULSE GENERATOR, PERMANENT, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC, INC. SENSIA IPG; PULSE GENERATOR, PERMANENT, IMPLANTABLE Back to Search Results
Model Number SEDR01
Medical Device Problem Code Battery Problem (2885)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/24/2015
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the implantable pulse generator (ipg) "battery went dead" and the patient did not know about it.The ipg was explanted and replaced.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Product event summary: the device was returned for battery depletion, after 98.3 months.The device was received in a no telemetry no output condition.The device loss of telemetry and all functions were due to a severely depleted battery.The device was fully functional and operated under normal current drain when powered with an external supply.There was no evidence of a high current drain condition, and no hybrid anomalies were observed.The device passed the labeled longevity calculation.It was determined that the device met specifications for normal operation and normal battery depletion.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENSIA IPG
Common Device Name
PULSE GENERATOR, PERMANENT, IMPLANTABLE
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key16478665
Report Number2182208-2023-00607
Device Sequence Number16928035
Product Code NVZ
UDI-Device Identifier00613994177957
UDI-Public00613994177957
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
P980035
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2007
Device Explanted Year2015
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date09/28/2008
Device Model NumberSEDR01
Device Catalogue NumberSEDR01
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/10/2023
Supplement Date Received by Manufacturer07/19/2023
Initial Report FDA Received Date03/03/2023
Supplement Report FDA Received Date07/21/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/03/2007
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
4068-45 LEAD, 4092-52 LEAD
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age88 YR
Patient SexFemale
-
-