| Model Number |
RIGID SADDLE RING |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Pericardial Effusion (3271); Heart Block (4444)
|
| Date of Event |
02/08/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Event or Problem Description
|
|
Crd_985 - arb pmcf (b)(6).It was reported that on (b)(6) 2023, a 30mm rigid saddle ring was successfully implanted.On (b)(6) 2023, it was reported that the patient had a complete atrioventricular block and a permeant pacemaker was implanted.The patient was reported to be in stable condition.
|
| |
|
Additional Manufacturer Narrative
|
|
An event of atrioventricular (av) block and pericardial effusion was reported.A returned device assessment could not be performed as the device remains implanted and was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.Based on the information received, the cause of the reported incident could not be conclusively determined.There is no indication of a product quality issue with respect to labeling design or manufacturing of the device.Arrhythmia is a potential side effect of the procedure per the rigid saddle ring instructions for use.
|
| |
|
Event or Problem Description
|
|
Crd_985 - arb pmcf eu2624-285 (b)(4).It was reported that on 01 february 2023, a 30mm rigid saddle ring was successfully implanted.On (b)(6) 2023, it was reported that the patient had a complete atrioventricular block and a permeant pacemaker was implanted.On (b)(6) 2023, the patient had a pericardial effusion during rehabilitation.On (b)(6) 2023, the patient underwent re-operation.
|
| |
|