| Model Number |
142122-15 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Discomfort (2330); Restenosis (4576)
|
| Date of Event |
01/23/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
There was no reported device malfunction and the device was not returned for analysis.A review has been performed of the manufacturing certificate of conformance associated with this lot.There were no non-conformances or deviations which could be associated with this event.Based on the case information and related record review, a conclusive cause for the reported patient effects, and the relationship to the product, if any, cannot be determined.The reported patient effects of restenosis leading to intervention and leg pain/claudication are listed in the biomimics 3d instructions for use and are known patient effect of peripheral stenting procedures.There is no indication of a product quality issue with respect to the design, manufacture or labelling of the device.If further information regarding this event becomes available, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
The patient was treated as part of the mimics 3d usa post-market observational study.On the (b)(6) 2022, the patient was implanted with one biomimics 3d (bm3d) stent (a 7.0 x 125mm stent) to treat a denovo lesion of the superficial femoral artery (sfa) proximal third in the right leg.A contralateral approach was used and the lesion was pre-dilated with percutaneous transluminal angioplasty (pta) and atherectomy.The treated segment was post-dilated with pta.The site reported that on (b)(6) 2023, a restenosis of treated segment (target lesion) was identified.The patient was seen in the clinical office on (b)(6) 2023 with some right leg discomfort.Duplex ultrasound (dus) showed high-grade velocities in proximal and mid sfa with significant dropping flows distal to this and decreasing ankle brachial indexes (abis).A right leg angiogram was conducted on (b)(6) 2023.The event was successfully treated with medication and percutaneous intervention, which involved a pta / standard balloon angioplasty of the sfa proximal third and anterior tibial artery on(b)(6) 2023 with good results.It was reported as a target lesion revascularisation (tlr) / target vessel revascularisation (tvr).The event of restenosis was reported as possibly related to the device and not related to the procedure.It was reported as target lesion related.The patient outcome was reported as resolved/recovered and the device remains implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
There was no reported device malfunction and the device was not returned for analysis.A review has been performed of the manufacturing certificate of conformance associated with this lot.There were no non-conformances or deviations which could be associated with this event.Based on the case information and related record review, a conclusive cause for the reported patient effects, and the relationship to the product, if any, cannot be determined.The reported patient effects of restenosis leading to intervention and leg pain/claudication are listed in the biomimics 3d instructions for use and are known patient effects of peripheral stenting procedures.There is no indication of a product quality issue with respect to the design, manufacture or labelling of the device.If further information regarding this event becomes available, a follow-up report will be submitted.Sections b.3.And b.5.Were updated to reflect the date of onset and updated distal segment details, sections d.3.And g.1.Were updated to reflect veryan's address, sections g.6.And h.2.Were updated to reflect the type of report (follow-up 01) and the reason.Section h.11.Was updated to reflect the sections changed in this report.
|
| |
|
Event or Problem Description
|
|
The patient was treated as part of the mimics 3d usa post-market observational study.On (b)(6)2022, the patient was implanted with one biomimics 3d (bm3d) stent (a 7.0 x 125 mm stent) to treat a denovo lesion of the superficial femoral artery (sfa) proximal third in the right leg.A contralateral approach was used and the lesion was prepared using pre-dilation with percutaneous transluminal angioplasty (pta) balloon and atherectomy.The treated segment was post-dilated with pta.The site reported that on (b)(6)2023, a restenosis of treated segment (target lesion) was identified.The patient was seen in the clinical office on (b)(6)2023 with some right leg discomfort.Duplex ultrasound (dus) showed high-grade velocities in proximal and mid sfa with significant dropping flows distal to this and decreasing ankle brachial indexes (abis).A right leg angiogram was conducted on (b)(6)2023.The event was successfully treated with medication and percutaneous intervention, which involved a pta / standard balloon angioplasty of the sfa proximal third and anterior tibial (at) distal third artery on (b)(6)2023 with good results.It was reported as a target lesion/vessel revascularisation (tlr/tvr).The event of restenosis was reported as possibly related to the device and not related to the procedure.It was reported as target lesion related.The patient outcome was reported as resolved/recovered and the device remains implanted.Additional information received on 14-nov-24 by veryan updated the date of onset of the event to 23-jan-23 and also updated the distal segment treated during the intervention to the at distal third.
|
| |
|