ROCHE DIAGNOSTICS TRIGLYCERIDES; TRIGLYCERIDES ASSAY Back to Search Results
Catalog Number 08058687190
Medical Device Problem Codes High Test Results (2457); Non Reproducible Results (4029)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/03/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The customer performed reagent carryover check and precision check and both checks passed.A general reagent problem was ruled out because qc results were acceptable.The investigation determined the issue is consistent with incorrect pre-analytic sample handling.No product problem was found.
 
Event or Problem Description
We received an allegation of questionable results for 1 patient's sample tested with the trigl assay on cobas pro c 503 analytical unit serial number (b)(4).On (b)(6) 2023 at 8:27 a.M.The customer ran 1 patient's sample and initially resulted in trigl value of 266 mg/dl.The result was reported to the physician but it did not correlate with the patient's clinical picture so he asked to retest the sample again.On the same day at 10:45 a.M.The patient's sample was repeated from the same primary tube and on the same analyzer and resulted in trigl value of 143 mg/dl.Later in the day, the patient's sample was repeated but this time on another c503 module, and resulted in trigl value of 144 mg/dl.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIGLYCERIDES
Common Device Name
TRIGLYCERIDES ASSAY
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
na
indianapolis, IN 46250
3174767531
MDR Report Key16479853
Report Number1823260-2023-00644
Device Sequence Number16931615
Product Code CDT
UDI-Device Identifier07613336121542
UDI-Public07613336121542
Combination Product (Y/N)N
Initial Reporter CountryTH
PMA/510(K) Number
K951595
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 03/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2023
Device Catalogue Number08058687190
Device Lot Number67754101
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/03/2023
Initial Report FDA Received Date03/03/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-