|
This report has also been submitted on importer medwatch #2429304 - 2023 - 00029.The suspect device was returned for evaluation without the clip.A visual inspection on the received condition was performed.The evaluator was unable to test the physical device due to the clip from the distal end on the sheath was missing as the clip was not returned for evaluation.However, the j-hook was noted intact from the distal end side.There was no sign of damages/kinks/dents found with the coil sheath.The slider movement on the handle side that opens and detaches the clips was deemed normal, while at the same time the tube sheath and coil sheath were coiled with no signs of restriction on the slider movement.The yellow joint below the handle area was at the expected location.The user reported problem could not be confirmed.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or when additional information becomes available.
|
|
An olympus representative reported to olympus in behalf of the customer that during an unknown procedure using this single use repositionable clip, the clip did not reopen after grabbing the tissue and caused more gastrointestinal bleeding.The only way to remove it, as reported was to rip it off.The bleeding was stopped using a competitor clip.The procedure was delayed for fifteen (15) minutes.No additional intervention was done other than more hemostasis that was needed.There were no reports of further patient or user harm associated with this event.
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.Correction to e2, e3, and d9 for information inadvertently left out.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Olympus confirmed "deploy prematurely" is reproduced when the clip arm is opened and closed using a method not recommended/included in the instructions for use (performed on lot 24k - 2yk and reproduced at 24k).In addition, the dimensions of the available lots (24k to 2yk) were checked to see if the occurrence of this phenomenon was caused by the dimensions of the parts.The dimensions of the hooks and clip arm related to this phenomenon were within the standard values.However, in 24k, the clearance between the hook and claw varied toward a smaller value compared to other lots."deploy prematurely" is more likely to occur when clearances are smaller.Based on the results of the investigation, an association between adverse events and the device could not be ruled out.It is likely the event occurred when an attempt was made to release the clip by pushing the slider, the convex area of the hook was facing down.Therefore, the hook did not detach under its own weight and the clip did not detach from the product.The hook may then have been released from the hook and the clip released by the time the product was removed from the scope.It is also possible that the slider has been pulled back to some degree and the clip is now ready for release.However, the exact cause could not be determined.The event can be detected and prevented by following the instructions for use which state: "operation of this instrument is based on the assumption that open surgery is possible as an emergency measure if the clip cannot be detached from the instrument or if any other unexpected circumstance takes place.In this case, refer to chapter 14, ¿emergency treatment¿.Keep the insertion portion of the instrument that extends from the biopsy valve and the insertion portion of the endoscope as straight as possible.Do not perform clipping with the insertion section being coiled.Otherwise, the force exerted on the control section may not convey to the clip adequately during clipping.This could result in reduced performance, making it impossible to close the clip or detach it from the coil sheath after clipping.Do not withdraw the instrument forcibly or change the angulation of the endoscope when the clip is grasping the tissue and is not released.Doing so may tear tissue inside the body cavity, resulting in patient injury, such as perforation, bleeding, or mucous membrane damage.Do not angulate the bending section of the endoscope or withdraw the instrument from the endoscope while the clip is grasping the tissue before being released.This could cause patient injury, such as perforation, bleeding, or mucous membrane damage.Do not try to forcibly withdraw the instrument from the endoscope if the clip cannot be detached from the instrument.Forcibly withdrawing the instrument could cause patient injury such as perforation, bleeding, or mucous membrane damage.Should the slightest irregularity and/or breakage of the parts be suspected, withdraw the instrument from the endoscope immediately according to the steps of ¿withdrawing the instrument from the endoscope¿ on page 10.Otherwise, perforation, bleeding, or mucous membrane damage may result." olympus will continue to monitor field performance for this device.
|