• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SOFRADIM PRODUCTION SAS PROGRIP; MESH, SURGICAL, POLYMERIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SOFRADIM PRODUCTION SAS PROGRIP; MESH, SURGICAL, POLYMERIC Back to Search Results
Model Number LPG1309AR
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 01/31/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during a hernia laparoscopic procedure, the mesh was torn after insertion into the patient.To resolve the issue, a new product was replaced in order to complete the case.
 
Additional Manufacturer Narrative
Additional information: d9, g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that the mesh tore.The reported issue could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRIP
Common Device Name
MESH, SURGICAL, POLYMERIC
Manufacturer (Section D)
SOFRADIM PRODUCTION SAS
116 avenue du formans
trevoux 01600
FR  01600
Manufacturer (Section G)
SOFRADIM PRODUCTION SAS
116 avenue du formans
trevoux 01600
FR   01600
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key16481917
Report Number9615742-2023-00291
Device Sequence Number12045479
Product Code FTL
UDI-Device Identifier10884521513198
UDI-Public10884521513198
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K143386
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2025
Device Model NumberLPG1309AR
Device Catalogue NumberLPG1309AR
Device Lot NumberPWB1232X
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/16/2023
Supplement Date Received by Manufacturer04/06/2023
Initial Report FDA Received Date03/03/2023
Supplement Report FDA Received Date04/18/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/24/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexMale
-
-