| Model Number |
ABGKT3CC23G-REG |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/02/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
There is potential contamination to patient and clinician.This would cause for the process to be repeated.
|
| |
|
Event or Problem Description
|
|
Leaking from the needle. .
|
| |
|
Additional Manufacturer Narrative
|
|
There is potential contamination to patient and clinician.This would cause for the process to be repeated.Summary: initial complaint was confirmed that blood could leak from needle of abg devices.Attempts to replicate failure mode indicated that dropping the device or pushing the rod/plunger assembly could cause fluid to leak from the syringe/needle assembly.Further conversations with customer indicated the customer does not prefer the device supplied by sunmed and mishandling of the device may be occurring as a cause of the leaks.Ra: rma-20036 risk id r78: clinician exposure to blood; s=8.
|
| |
|
Event or Problem Description
|
|
Leaking from the needle. .
|
| |
|