• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AOMORI OLYMPUS CO., LTD. DISPOSABLE BIOPSY FORCEPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AOMORI OLYMPUS CO., LTD. DISPOSABLE BIOPSY FORCEPS Back to Search Results
Model Number FB-230U
Medical Device Problem Code Dull, Blunt (2407)
Health Effect - Clinical Codes Hemorrhage/Blood Loss/Bleeding (1888); Unspecified Tissue Injury (4559)
Date of Event 01/16/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The investigation is ongoing and follow up with the user facility is currently being performed.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Event or Problem Description
A user facility reported to olympus that while using a disposable biopsy forceps to obtain a tissue sample during a colonoscopy with biopsy procedure, the surgeon complained about the jaw not being sharp.This caused a bad cutting border.Additionally it tore tissue in the submucosa in the colon and bleeding occurred.There was a delay of an unknown duration that occurred while the surgeon needed to observe the bleeding area, but no additional device was needed to stop the bleeding.The amount of blood loss was unknown.Additional information regarding interventions and outcome was requested multiple times without response.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it was determined that an abnormality of the device was not the cause of the event.As reported, the nurse may have assumed that the device (model # fb-230u) was not sharp enough and may have torn the patient's tissue, causing the bleed.However, the root cause of the reported event could not be identified.The event can be detected/prevented by following the instructions for use (ifu) which state: ¿do not force the distal end of the insertion portion against body cavity tissue.Doing so could cause patient injuries, such as perforation, bleeding, or mucous membrane damage.¿ ¿biopsy may cause bleeding.If you take a large sample or press the instrument¿s distal end against tissue excessively, the risk of bleeding will increase.Perform biopsy on minimum necessary spots only.¿ this supplemental report includes a correction to e3 to provide information that was inadvertently not included in the initial medwatch.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DISPOSABLE BIOPSY FORCEPS
Common Device Name
DISPOSABLE BIOPSY FORCEPS
Manufacturer (Section D)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori 036-0 357
JA  036-0357
Manufacturer (Section G)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key16482134
Report Number9614641-2023-00312
Device Sequence Number14768227
Product Code FCL
UDI-Device Identifier04953170031120
UDI-Public04953170031120
Combination Product (Y/N)N
Initial Reporter CountryTH
PMA/510(K) Number
CLASS1-EXMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 05/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberFB-230U
Device Lot Number26T
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/01/2023
Supplement Date Received by Manufacturer05/19/2023
Initial Report FDA Received Date03/03/2023
Supplement Report FDA Received Date05/26/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-