| Model Number |
SCPX107MH |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Deformity/ Disfigurement (2360)
|
| Date of Event |
01/01/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: material rupture mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a female patient underwent breast prosthesis implantation surgery with a 275cc mentor cpx4 plus, smooth, medium height on an unspecified breast and suffered breast implant deflation, post-operatively.As a result, the patient underwent breast implant removal surgery on (b)(6) 2023.
|
| |
|
Additional Manufacturer Narrative
|
|
On march 30, 2023, the mentor failure analysis lab received the device for evaluation.On april 18, 2023, the product investigation was completed.Device investigation summary: the product was returned to mentor for evaluation.Mentor conducted a visual inspection, leak testing, and microscopic examination of the returned tissue expander.Visual analysis of the returned sample revealed that no damage or anomalies were observed on the cpx4 plus sm te mh 275cc tissue expander.Leak testing was performed, in accordance with mentor procedures, and a tear was noted on the anterior aspect measuring approximately 0.1 cm.Microscopic examination was performed on the edges of the rupture, and parallel striations were found in the whole area of the tear.Parallel striations are consistent with markings made by a sharp object perforating the tissue expander.Based on the information currently available, a microscopic examination of the returned product indicates that the tissue expander could have been damaged during or subsequent to implantation.The product insert data sheet cautions to not allow cautery devices or sharp instruments, such as scalpels, suture needles, hypodermic needles, hemostats, adson forceps, or scissors to contact the device during the implantation or other surgical procedures.As part of mentor¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Capa activity was not required.
|
| |
|
Search Alerts/Recalls
|
|
|