| Model Number |
20E |
| Medical Device Problem Code |
Failure to Sense (1559)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/14/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer contacted stryker to report a non-critical issue with their device.During evaluation it was found that the device did not recognize the defibrillation therapy cable.In this state, the device may not be able to deliver defibrillation therapy if needed.There was no patient involvement reported during the event.
|
| |
|
Additional Manufacturer Narrative
|
|
Stryker evaluated the customer's device and was able to verify and duplicate the reported issue.The cause of the reported issue was isolated to the w09 cable and replaced it to resolve the issue.After completing other unrelated repairs, proper device operation was observed through functional and performance testing and the device returned to the customer for use.
|
| |
|