| Catalog Number |
20374 |
| Medical Device Problem Codes |
False Positive Result (1227); No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/14/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Reported false positive sars result for 1 patient.It is unknown if the patient was symptomatic.The customer communicated the result was confirmed negative by molecular (pcr testing).It was reported that the customer had been using remel m4-rt utm as well as third party swabs (off-label use) to run the tests.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation conclusion: a review of the package insert (pi) was conducted for clarity of instructions.No issues were found.The customer's reported problem was related to a deviation from the instructions called out in the pi.Root cause: customer procedural error.Source: phone.
|
| |
|
Search Alerts/Recalls
|