• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: A.I.D.D LONGFORD MAGNESIUM; PHOTOMETRIC METHOD, MAGNESIUM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

A.I.D.D LONGFORD MAGNESIUM; PHOTOMETRIC METHOD, MAGNESIUM Back to Search Results
Model Number 08P1934
Medical Device Problem Code Low Test Results (2458)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/14/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.
 
Event or Problem Description
The customer observed a falsely decreased alinity c magnesium result for one patient.The result was not reported out.The sample was repeated and the result was higher.The following data was provided processed on 14feb2023: sid (b)(6) initial result = 0.48 repeat result = 1.12.Reference range = 0.66 to 1.07 mmol/l no negative impact to patient management was reported.
 
Additional Manufacturer Narrative
The complaint investigation included a review of data and information provided by the customer, search for similar complaints, ticket trending review, labeling review, and device history record review.Return testing was not completed as returns were not available.A search for similar complaints and ticket trending review did not identify any trends or issues with the complaint lot number or current complaint issue.Device history record review did not identify any non-conformances or deviations related to the complaint lot and complaint issue.The historical performance of reagent lot 43489ud00 was evaluated using data from customers worldwide.The patient data was analyzed and compared to an established control limit.This evaluation indicated that the patient median result for lot 43489ud00 is within the established control limits.Therefore, no unusual reagent lot performance was identified for lot 43489ud00.Labeling was reviewed and was found to adequately address the issue under review.Based on the investigation, no systemic issue or deficiency was identified for the alinity c magnesium reagent, lot 43489ud00.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAGNESIUM
Common Device Name
PHOTOMETRIC METHOD, MAGNESIUM
Manufacturer (Section D)
A.I.D.D LONGFORD
lisnamuck
co. longford
longford N39E9 32
EI  N39E932
Manufacturer (Section G)
A.I.D.D LONGFORD
lisnamuck
co. longford
longford N39E9 32
EI   N39E932
Manufacturer Contact
siobhan wright
lisnamuck
post market surveillance
longford N39 E-932
EI   N39 E932
433331157
MDR Report Key16483458
Report Number3005094123-2023-00061
Device Sequence Number20149446
Product Code JGJ
UDI-Device Identifier00380740176532
UDI-Public00380740176532
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K181748
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/17/2023
Device Model Number08P1934
Device Catalogue Number08P19-34
Device Lot Number43489UD00
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/15/2023
Supplement Date Received by Manufacturer04/07/2023
Initial Report FDA Received Date03/04/2023
Supplement Report FDA Received Date04/12/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/12/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ALNTY C PROCESSING MODU, 03R67-01, AC02346; ALNTY C PROCESSING MODU, 03R67-01, AC02346
Patient SexUnknown
-
-