| Model Number |
0227-6105 |
| Medical Device Problem Codes |
Break (1069); Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/08/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Once the investigation has been completed any additional information will be reported in a supplemental report.
|
| |
|
Event or Problem Description
|
|
It was reported that an im flexible reamer was broken during surgery which resulted in a 20 minute surgical delay.There was no debris.
|
| |
|
Additional Manufacturer Narrative
|
|
Please note the correction to d4 lot #.The reported event could be confirmed since the device was returned and matches the alleged failure.Device inspection revealed the following: from the visual inspection of the returned im reamer where the spiral was also found to be broken from the point where it is welded with the adapter.Around that region, a burn mark can be seen which is a sign of excessive restricted rotation up to a point of heat generation.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.Based on the investigation, the root cause was attributed to user related as breakage was caused due to application of high torsional load.If any additional information is provided, the investigation will be reassessed.
|
| |
|
Event or Problem Description
|
|
It was reported that an im flexible reamer was broken during surgery which resulted in a 20 minute surgical delay.There was no debris.
|
| |
|