| Model Number |
ECS21A |
| Medical Device Problem Code |
Failure to Fire (2610)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/08/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4) batch # unk.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be field as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that during the esophagus surgery, could not fire.Changed the new one to complete surgery.There was no patient consequence reported.No additional information can be provided.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 03/21/2023.D4: batch # 946a07 investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the ecs21a device arrived with no apparent damage with an anvil inside of a plastic bag.The breakaway washer was present, uncut and the device was fully loaded with staples.Further analysis of the device showed that the trocar was damaged no allowing the anvil to properly assemble on the device.No functional test was performed due to the condition of the trocar.This defect has been correlated to a manufacturing process.A manufacturing record evaluation was performed for the finished device batch number 946a07, and no non-conformances were identified.
|
| |
|
Search Alerts/Recalls
|