U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 BROACH CORAIL AMT 12; HIP INSTRUMENTS : BROACHES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 BROACH CORAIL AMT 12; HIP INSTRUMENTS : BROACHES Back to Search Results
Model Number L20412
Medical Device Problem Codes Device-Device Incompatibility (2919); Material Deformation (2976); Naturally Worn (2988)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 02/20/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).Component code: appropriate term/code not available (g07002) used to capture no findings available.Investigation summary: no device was received for examination, therefore the reported event could not be confirmed.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the size 12 broach had a dent in it at the top part which connects to the neck trials and wouldn¿t allow the necks to fully seat down on it.This means a proper trial is unable to be carried out with this size.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BROACH CORAIL AMT 12
Common Device Name
HIP INSTRUMENTS : BROACHES
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY IRELAND 9616671
loughbeg, ringaskiddy co.
cork
EI  
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key16486121
Report Number1818910-2023-05056
Device Sequence Number12237135
Product Code HTQ
UDI-Device Identifier10603295258148
UDI-Public10603295258148
Combination Product (Y/N)N
Initial Reporter CountryUK
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 03/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberL20412
Device Catalogue NumberL20412
Device Lot Number1893680
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/20/2023
Initial Report FDA Received Date03/06/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-