CORDIS US. CORP SABER 2MM15CM 150; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL Back to Search Results
Model Number 48002015X
Medical Device Problem Code Burst Container or Vessel (1074)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/10/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional information is pending and will be submitted within 30 days upon receipt.
 
Event or Problem Description
As reported, the balloon of a 2mm x 15cm 150cm saber percutaneous transluminal angioplasty (pta) balloon catheter was inflated normally but had ruptured prior to reaching the pressure.There was no reported injury to the patient.Additional information was requested but was not provided.The device was not returned as expected.
 
Additional Manufacturer Narrative
Complaint conclusion: as reported, the balloon of a 2mm x 15cm 150cm saber percutaneous transluminal angioplasty (pta) balloon catheter was inflated normally but had ruptured prior to reaching the pressure.There was no reported injury to the patient.Additional information was requested but was not provided.The product was not returned for analysis as expected.A product history record (phr) review of lot 82234635 revealed no anomalies or non-conformances during the manufacturing and inspection processes that can be associated with the reported event.The reported ¿balloon burst-at/below rbp¿ could not be confirmed as the device was not returned for analysis.The exact cause could not be determined.Vessel characteristics, although unknown and procedural factors likely contributed to the reported event.The procedure performed was below the knee, these vessels may be tortuous in nature due to location; however, lesion characteristics were not provided.Therefore, without the return of the device for analysis it is difficult to draw a clinical conclusion between the device and the event reported.According to the warnings in the safety information in the instructions for use ¿prior to angioplasty, the catheter should be examined to verify functionality and integrity and ensure that its size and shape are suitable for the specific procedure for which it is to be used.Do not use if product damage is suspected or evident.To reduce the potential for vessel damage or the risk of dislodgement of particles it is very important that the inflated diameter of the balloon should approximate the diameter of the vessel just proximal and distal to the lesion.The balloon dimensions are printed on the product label.The compliance table incorporated with the product shows how balloon diameter increases as pressure increases.Do not exceed the rated burst pressure recommended on the label.The rated burst pressure is based on the results of in vitro testing.At least 99.9% of the balloons (with a 95% confidence) will not burst at or below their rated burst pressure.Use of a pressure monitoring device is recommended to prevent over-pressurization.Pressure in excess of the rated burst pressure can cause balloon rupture and potential inability to withdraw the catheter through the introducer sheath.Balloon rupture can cause vessel damage and the need for additional intervention.Use only the recommended balloon inflation medium (a 50/50 mixture by volume of contrast medium and normal saline).Never use air or any gaseous medium to inflate the balloon.¿ neither the phr nor the information available suggests a design or manufacturing related cause for the reported event.Therefore, no corrective or preventive action will be taken at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SABER 2MM15CM 150
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
CORDIS US. CORP
14201 nw 60 avenue
miami lakes FL 33014
Manufacturer (Section G)
CORDIS US. CORP
14201 nw 60 avenue
miami lakes FL 33014
Manufacturer Contact
karla castro
14201 nw 60 avenue
miami lakes, FL 33014
7863138372
MDR Report Key16486586
Report Number9616099-2023-06300
Device Sequence Number12452226
Product Code LIT
UDI-Device Identifier20705032069202
UDI-Public(01)20705032069202(17)241031(10)82234635
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K971010
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2024
Device Model Number48002015X
Device Catalogue Number48002015X
Device Lot Number82234635
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/09/2023
Supplement Date Received by Manufacturer03/15/2023
Initial Report FDA Received Date03/06/2023
Supplement Report FDA Received Date04/10/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/23/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-