• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMO PUERTO RICO MFG. INC. TECNIS SIMPLICITY; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMO PUERTO RICO MFG. INC. TECNIS SIMPLICITY; INTRAOCULAR LENS Back to Search Results
Model Number DCB00
Medical Device Problem Codes Break (1069); Crack (1135)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/08/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that an intraocular lens (iol) cartridge tears at the end while inserting the lens into the patient's operative eye.This occurs as the surgeon screws the lens through into the eye.The implant also has some small little defects on it.The surgeon sees abnormalities on the lens.Nurse assured she filled the cartridge with balanced salt solution (bss) and sat for more than 3-5 minutes.
 
Additional Manufacturer Narrative
Model number: a complete model is unknown, as the serial number was not provided.Catalog number: a complete catalog number is unknown, as the serial number was not provided.If implanted, give date: unknown/not reported; it is unknown if the iol had any patient contact or if it was implanted.If explanted, give date: unknown/not reported; it is unknown if the iol had any patient contact or if it was implanted.Initial reporter telephone number: (b)(6).The device was not returned for evaluation; therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.An attempt has been made to obtain missing information; however, no definitive response has been received.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
Additional info: further information was provided and reported the date of event and the serial number was provided.The surgeon also reported he still used the iol.No other information was provided.The following sections have been updated accordingly: section b3: date of event: feb 8, 2023.Section d4: model number: dcb00.Section d4: catalog number: dcb0000135.Section d4: serial number: (b)(6).Section d4: expiration date: nov 20, 2025.Section d4: udi number: (b)(4).Section d6a: if implanted, give date: (b)(6) 2023.Section d6b: if explanted; give date: n/a (not applicable).The lens remains implanted.Section h4: device manufacture date: nov 20, 2022.Device evaluation: the customer provided a photo and a visual inspection of the photo revealed a stress mark on the cartridge leading to the cartridge tip.The stress mark identified was observed to be within specification because no lens was stuck inside the cartridge, and no cartridge or cartridge tip crack was identified.The cartridge tip was also slightly deformed.No photo of the complaint lens was received; therefore, the complaint issue of lens damage could not be confirmed.No further product evaluation could be performed.The complaint issue " cartridge tip cracked/damaged was identified during product evaluation; however, based on the complaint investigation results the complaint issue could not be confirmed to be related to the manufacturing or design process.The other observed issue found during the product evaluation could not be confirmed to be related to a manufacturing or design issue.Conclusion: based on the complaint investigation results, the product was released within specifications and a relationship between the device and the reported incident could not be determined.As a result of the investigation there is no product quality deficiency, and the reported issue could not be identified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS SIMPLICITY
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key16487166
Report Number3012236936-2023-00582
Device Sequence Number5899897
Product Code HQL
UDI-Device Identifier05050474635951
UDI-Public(01)05050474635951(17)251120
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/29/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/20/2025
Device Model NumberDCB00
Device Catalogue NumberDCB0000135
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/08/2023
Supplement Date Received by Manufacturer04/05/2023
Initial Report FDA Received Date03/06/2023
Supplement Report FDA Received Date04/29/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/20/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-