• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: JJGC S.A. CM DRIVE IMPL TI 4.3X16; ENDOSSEOUS DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

JJGC S.A. CM DRIVE IMPL TI 4.3X16; ENDOSSEOUS DENTAL IMPLANT Back to Search Results
Model Number 109.629
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Pain (1994)
Date of Event 02/22/2021
Type of Reportable Event Serious Injury
Event or Problem Description
The clinician reports the implant was inserted (b)(6) 2015 in ada 20.On (b)(6) 2021, fracture of the implant was verified.No product was returned to the manufacturer.At the event the patient experienced: pain.No further patient complications were reported.
 
Additional Manufacturer Narrative
The batch number could be verified.Our manufacturing q-system assures, that production and process controls are in place to ensure that batches confirm to the applicable specifications before they are distributed.The fracture of a dental implant is a known long-term complication of endosseous dental implants.The manufacturers trend analysis confirms that the reported implant fracture rate associated with its dental implants is low and remains consistent with the experience as documented in the literature.There are different reasons why implants break and this problem has been illuminated by various authors (e.G.Maeglin 1988, beumer 1989, tetsch 1991, worthington 1992).Beumer and lewis 1989 report the following causes: a) production / design flaws b) inadequate fitting of the suprastructure c) occlusal overload d) bruxism e) bone resorption around the implants f) insufficient axle alignment of the implants g) trauma h) biomechanical causes i) design of the suprastructure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CM DRIVE IMPL TI 4.3X16
Common Device Name
ENDOSSEOUS DENTAL IMPLANT
Manufacturer (Section D)
JJGC S.A.
av. juscelino kubitschek de
oliveira, 3291 cic
curitiba PR 81270 -200
BR  81270-200
Manufacturer (Section G)
JJGC S.A.
av. juscelino kubitschek de
oliveira, 3291 cic
curitiba PR 81270 -200
BR   81270-200
Manufacturer Contact
jennifer jackson
60 minuteman road
andover, MA 01810
9787472509
MDR Report Key16489977
Report Number3008261720-2023-005366
Device Sequence Number6787616
Product Code DZE
UDI-Device Identifier07898237568775
UDI-Public07898237568775
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K123022
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 03/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/01/2019
Device Model Number109.629
Device Catalogue Number109.629
Device Lot Number800136397
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Distributor Facility Aware Date03/06/2023
Event Location Other
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date03/06/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/01/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-